Characteristics and Clinical Effects of Myopotential Signals in a Unipolar DDD Pacemaker Population

Myopotential signals were recorded from atrial and ventricular leads during isometric exercise in 25 patients who had chronically implanted dual chamber pacemakers using the electrogram telemetry capability of the pacemakers. Average electrogram amplitude on the atrial channel was 0.92 mV (range 0.3...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 1986-11, Vol.9 (6), p.1019-1025
Hauptverfasser: ZIMMERN, SAMUEL H., CLARK, MARTHA F., AUSTIN, W. KENNETH, FEDOR, JOHN M., GALLAGHER, JOHN J., SVENSON, ROBERT H., DUNCAN, JAMES L.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Myopotential signals were recorded from atrial and ventricular leads during isometric exercise in 25 patients who had chronically implanted dual chamber pacemakers using the electrogram telemetry capability of the pacemakers. Average electrogram amplitude on the atrial channel was 0.92 mV (range 0.3 to 1.9) and on the ventricular channel was 0.98 mV (range 0.3 to 2.2); the difference was not significant. There was a strong correlation (R = 0.82) between the amplitude of myopotentials on the atrial and ventricular leads for individual patients. Myopotential sensing caused ventricular output inhibition in two patients (8%) and ventricular tracking in sixteen patients (64%). Pacemaker reprogramming abolished ventricular myopotential inhibition in all patients and stopped ventricular myopotential tracking in seven patients. We conclude that myopotentials can be analyzed and their effects ameliorated by a multiprogrammable pacemaker with electrogram telemetry capability.
ISSN:0147-8389
1540-8159
DOI:10.1111/j.1540-8159.1986.tb06664.x