Combination of Potentiometry and Resonance Raman Spectroscopy for the Analysis of a Redox Protein

This paper describes apparatus and procedures for combining resonance Raman and optical absorption spectroscopies with potentiometry for the study of redox-active heme proteins. A specially designed anaerobic titration cell is described which allows for the laser excitation of the sample and the mon...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical biochemistry 1995, Vol.224 (1), p.309-314, Article 309
Hauptverfasser: Harmon, P.A., Hendler, R.W., Friauf, W.S., Levin, I.W.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This paper describes apparatus and procedures for combining resonance Raman and optical absorption spectroscopies with potentiometry for the study of redox-active heme proteins. A specially designed anaerobic titration cell is described which allows for the laser excitation of the sample and the monitoring of both Raman scattered light and directly transmitted light from an optical source. New procedures for utilization of A/D and D/A converters on a standard I/O computer card are described, which allow for computer-controlled potentiometry and coulometry. The system was tested with cytochrome c, a well-characterized respiratory protein. The correct values for the midpoint potential and electron number of the Nernst equation were obtained both by the optical absorption and resonance Raman measurements.
ISSN:0003-2697
1096-0309
DOI:10.1006/abio.1995.1045