Sampling Errors in pH and Blood Gas Analysis—An Evaluation of Three New Arterial Blood Samplers

We have tested the accuracy, acceptability and general performance of three recently-marketed samplers for arterial blood gas measurement (the Corning Arterial Blood Sampler, the Concord ‘Pulsator’ and the Sarstedt ‘Monovette’™). All three greatly reduce or eliminate the error of venous sampling, an...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annals of clinical biochemistry 1986-05, Vol.23 (3), p.329-333
Hauptverfasser: Hutchison, A S, Dryburgh, F J, Ralston, S H
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We have tested the accuracy, acceptability and general performance of three recently-marketed samplers for arterial blood gas measurement (the Corning Arterial Blood Sampler, the Concord ‘Pulsator’ and the Sarstedt ‘Monovette’™). All three greatly reduce or eliminate the error of venous sampling, and the Corning and Sarstedt samplers eliminate the risk of dilution of the sample by excess heparin solution. A positive bias in pO2 measurement, more marked at higher levels, was demonstrated with the Concord and Sarstedt samplers, and the latter carry a slightly increased risk of cross-infection. None of the samplers completely overcame potential sampling errors.
ISSN:0004-5632
1758-1001
DOI:10.1177/000456328602300314