Implantation of transvenous nonthoracotomy cardioverter-defibrillator systems in patients with permanent endocardial pacemakers
Among 177 patients in whom a nonthoracotomy approach was initially used to implant a cardioverter-defibrillator system, 11 (6%) patients also received a separately implanted permanent pacemaker. The main problem encountered in these patients were previously implanted unipolar pacemakers ( n = 3) and...
Gespeichert in:
Veröffentlicht in: | The American heart journal 1995, Vol.129 (1), p.45-53 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Among 177 patients in whom a nonthoracotomy approach was initially used to implant a cardioverter-defibrillator system, 11 (6%) patients also received a separately implanted permanent pacemaker. The main problem encountered in these patients were previously implanted unipolar pacemakers (
n = 3) and ventricular pacing leads positioned at the right ventricular apex, the latter interfering with optimal placement of the tripolar implantable cardioverter-defibrillator (ICD) lead (
n = 9). The approaches used to solve these problems were individualized and included placement of the ICD sensing lead at the right ventricular outflow tract (
n = 3), initial placement (
n = 1) or subsequent repositioning (
n = 2) of the right ventricular pacing lead at the outflow tract, upgrade from unipolar to bipolar systems (
n = 2), reprogramming from the DDD to AAI mode (
n = 2), inactivation of the pacemaker (
n = 1), and simultaneous placement of a single-chamber atrial pacemaker with the ICD lead (
n = 2). These revisions fulfilled the pacing needs in each patient and prevented unfavorable sensing interaction between the two systems. |
---|---|