Fracture-embolization of Duromedics valve prosthesis and microscopic uncommon lesions

We report a sudden leaflet fracture of a Duromedics mitral valve 6 years after implantation. The patient had cardiogenic shock and complained of asthenia, orthopnea, and tachycardia. Transesophageal echocardiography showed the lack of one leaflet of the prosthesis and regurgitation. An emergency mit...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Annals of thoracic surgery 2001-04, Vol.71 (4), p.1366-1369
Hauptverfasser: Tatou, Etienne, Saleh, Mountajab, Eicher, Jean Christophe, Brenot, Roger, David, Michel
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report a sudden leaflet fracture of a Duromedics mitral valve 6 years after implantation. The patient had cardiogenic shock and complained of asthenia, orthopnea, and tachycardia. Transesophageal echocardiography showed the lack of one leaflet of the prosthesis and regurgitation. An emergency mitral replacement was successfully performed. Angiographic computed tomography scan localized the sequestrum that embolized the common iliac arteries. Examination of the deficient prosthesis showed multiple lesions and, in particular, a subsurface lesion that may be characteristic of carbon pyrolytic valves.
ISSN:0003-4975
1552-6259
DOI:10.1016/S0003-4975(00)02368-7