Measurement of oxygen uptake in newborn infants during assisted and spontaneous ventilation
Measurements of oxygen uptake (V O 2 ) and CO 2 output (V CO 2 ) are useful in critically ill patients. However, V O 2 is not routinely measured in intensive care during mechanical ventilation (MV) especially in premature newborns. The present study describes a new method of measuring V O 2 and V O...
Gespeichert in:
Veröffentlicht in: | Respiration physiology 1994-10, Vol.98 (2), p.227-239 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Measurements of oxygen uptake (V
O
2
) and CO
2 output (V
CO
2
) are useful in critically ill patients. However, V
O
2
is not routinely measured in intensive care during mechanical ventilation (MV) especially in premature newborns. The present study describes a new method of measuring V
O
2
and V
O
2
using a double open circuit which accounts for gas leaks around the uncuffed tracheal tube. The accuracy of the method was assessed with N
2 and CO
2 infusion. In case of leaks, V
O
2
and V
O
2
measurement was significantly underestimated by the simple circuit method. This underestimation was not present with double circuit method. Five preterm newborns were studied. V
O
2
adn V
CO
2
using the double open circuit were compared with the classic simple circuit. During MV, the mean underestimation assessed by the difference between simple and double circuit measurement was −12% (range from 0 to −29%) for V
O
2
and −14%(range 0 to −26%) for V
CO
2
. |
---|---|
ISSN: | 0034-5687 |
DOI: | 10.1016/0034-5687(94)00056-5 |