Enzyme electrode for the measurement of lactate in undiluted blood

An enzyme electrode is described based upon the enzyme lactate oxidase (EC 1.1.3.2) coupled to an H 2O 2 sensor. Use of an organosilane-treated microporous membrane over the enzyme layer, allows responses linear to 18 mmol/l l-lactate with response times of 1–3 min, whilst giving independence from s...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinica chimica acta 1986-06, Vol.157 (2), p.191-198
Hauptverfasser: Mullen, W.H., Churchouse, S.J., Keedy, F.H., Vadgama, P.M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An enzyme electrode is described based upon the enzyme lactate oxidase (EC 1.1.3.2) coupled to an H 2O 2 sensor. Use of an organosilane-treated microporous membrane over the enzyme layer, allows responses linear to 18 mmol/l l-lactate with response times of 1–3 min, whilst giving independence from stirring. The method is suitable for the bedside analysis of a single drop of undiluted whole blood. Correlation with a routine method gave y = 0.96 x + 0.11, r = 0.996, n = 46.