Enzyme electrode for the measurement of lactate in undiluted blood
An enzyme electrode is described based upon the enzyme lactate oxidase (EC 1.1.3.2) coupled to an H 2O 2 sensor. Use of an organosilane-treated microporous membrane over the enzyme layer, allows responses linear to 18 mmol/l l-lactate with response times of 1–3 min, whilst giving independence from s...
Gespeichert in:
Veröffentlicht in: | Clinica chimica acta 1986-06, Vol.157 (2), p.191-198 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An enzyme electrode is described based upon the enzyme lactate oxidase (EC 1.1.3.2) coupled to an H
2O
2 sensor. Use of an organosilane-treated microporous membrane over the enzyme layer, allows responses linear to 18 mmol/l
l-lactate with response times of 1–3 min, whilst giving independence from stirring. The method is suitable for the bedside analysis of a single drop of undiluted whole blood. Correlation with a routine method gave
y = 0.96
x + 0.11,
r = 0.996,
n = 46. |
---|---|
ISSN: | 0009-8981 1873-3492 |
DOI: | 10.1016/0009-8981(86)90225-1 |