Fourteen years' experience with the Bjork-Shiley tilting disc prosthesis

Between 1970 and 1984, 1,574 Björk-Shiley valve prostheses have been implanted in 1,171 patients in Glasgow. Between 1970 and 1980, 1,023 standard disc prostheses were implanted in 729 patients: 184 had aortic valve replacement, 323 mitral, and 222 multiple valve replacement. Between 1980 and 1984,...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of thoracic and cardiovascular surgery 1986-03, Vol.91 (3), p.350-361
Hauptverfasser: Sethia, B, Turner, MA, Lewis, S, Rodger, RA, Bain, WH
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Between 1970 and 1984, 1,574 Björk-Shiley valve prostheses have been implanted in 1,171 patients in Glasgow. Between 1970 and 1980, 1,023 standard disc prostheses were implanted in 729 patients: 184 had aortic valve replacement, 323 mitral, and 222 multiple valve replacement. Between 1980 and 1984, 551 convexo-concave valves were implanted in 442 patients: 125 had aortic, 228 mitral, and 89 multiple valve replacement. Concomitant coronary artery bypass grafting was performed in 9.2% of the latter group as compared with 2.7% of the earlier group. The overall hospital (30 day) mortality for the group with the standard disc prosthesis was 10.4% and for patients with the convexo-concave prosthesis, 7.0%. Cumulative follow-up extends to 4,125 patient-years--standard disc group 3,378 patients-years, range 4 to 14 years (mean 5.2 years); convexo-concave group 747 patient-years, range 1.5 to 5.3 years (mean 1.9 years). The late mortality is 3.8% per patient-year--standard disc group 2.9% per patient-year and convexo-concave group 4.3% per patient year (no significant difference). Actuarial survival rate at 12 years for the whole group, excluding operative deaths, is 69.4% +/- 6.3%. The freedom from all valve-related complications at 12 years is 65.9% +/- 6.5%. there is no significant difference in the incidence of most major complications between the standard disc and convexo-concave prostheses apart from the occurrence of mitral valve thrombosis (p less than 0.05) in the standard disc group and outlet strut fracture (p less than 0.005) in the convexo-concave group. These results confirm the excellent long-term performance of the Björk-Shiley tilting disc prosthesis.
ISSN:0022-5223
1097-685X
DOI:10.1016/s0022-5223(19)36050-7