EVALUATION OF KEYED FILLERS ON TEC VAPORIZERS

Single and dual port versions of the keyed fillers fitted to TEC4 and TEC3 vaporizers, respectively, were examined. The single port filler performed satisfactorily, provided that it was correctly used and properly maintained. Design deficiencies in the dual-port keyed filler were found which account...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:British journal of anaesthesia : BJA 1986-03, Vol.58 (3), p.353-356
Hauptverfasser: RICHARDSON, W., CARTER, K.B.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Single and dual port versions of the keyed fillers fitted to TEC4 and TEC3 vaporizers, respectively, were examined. The single port filler performed satisfactorily, provided that it was correctly used and properly maintained. Design deficiencies in the dual-port keyed filler were found which accounted for many of the problems described by users. The keyed bottle adaptor could be improved by an increase in its length and the provision of two openings into the outer tube at the bottle ends.