Pacemaker Backup-Mode Reversion and Injury during Cardiac Surgery

Both electrocautery and internal defibrillation are routinely used in cardiac operations. The cases of 5 patients are presented in whom backup-mode reversion or injury of permanently implanted pulse generators occurred during a cardiac procedure. The theoretical explanations for backup-mode reversio...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Annals of thoracic surgery 1986-02, Vol.41 (2), p.155-157
Hauptverfasser: Lamas, Gervasio A., Antman, Elliott M., Gold, Jeffrey P., Braunwald, Nina S., Collins, John J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Both electrocautery and internal defibrillation are routinely used in cardiac operations. The cases of 5 patients are presented in whom backup-mode reversion or injury of permanently implanted pulse generators occurred during a cardiac procedure. The theoretical explanations for backup-mode reversion and generator or tissue injury are discussed, and recommendations are made for the management of patients with a pacemaker who are to undergo a cardiac operation.