Evaluation of a semi-automated 24-hour commercial system for identification of Enterobacteriaceae and other gram-negative bacteria
A semi-automated commercial system (ID 32 E, bioMérieux) for 24-hour identification of Enterobacteriaceae and other gram-negative fermentative and nonfermentative bacteria encountered in diagnostic microbiology was evaluated. Overall, the system correctly identified 506 (91.5%) of the 553 strains te...
Gespeichert in:
Veröffentlicht in: | European journal of clinical microbiology & infectious diseases 1994-05, Vol.13 (5), p.424-430 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A semi-automated commercial system (ID 32 E, bioMérieux) for 24-hour identification of Enterobacteriaceae and other gram-negative fermentative and nonfermentative bacteria encountered in diagnostic microbiology was evaluated. Overall, the system correctly identified 506 (91.5%) of the 553 strains tested, 94 (17.0%) strains requiring additional tests for complete identification. Six (1.1%) strains were misidentified and 33 (6.0%) strains were not identified. Eight (1.4%) strains were not present in the database and were misidentified or not identified. The system is a suitable alternative to existing systems for the identification of Enterobacteriaceae and other gram-negative bacteria frequently encountered in clinical samples. |
---|---|