Evaluation of a semi-automated 24-hour commercial system for identification of Enterobacteriaceae and other gram-negative bacteria

A semi-automated commercial system (ID 32 E, bioMérieux) for 24-hour identification of Enterobacteriaceae and other gram-negative fermentative and nonfermentative bacteria encountered in diagnostic microbiology was evaluated. Overall, the system correctly identified 506 (91.5%) of the 553 strains te...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European journal of clinical microbiology & infectious diseases 1994-05, Vol.13 (5), p.424-430
Hauptverfasser: MONNET, D, LAFAY, D, DESMONCEAUX, M, BOEUFGRAS, J. M, ALLARD, F, FRENEY, J
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A semi-automated commercial system (ID 32 E, bioMérieux) for 24-hour identification of Enterobacteriaceae and other gram-negative fermentative and nonfermentative bacteria encountered in diagnostic microbiology was evaluated. Overall, the system correctly identified 506 (91.5%) of the 553 strains tested, 94 (17.0%) strains requiring additional tests for complete identification. Six (1.1%) strains were misidentified and 33 (6.0%) strains were not identified. Eight (1.4%) strains were not present in the database and were misidentified or not identified. The system is a suitable alternative to existing systems for the identification of Enterobacteriaceae and other gram-negative bacteria frequently encountered in clinical samples.
ISSN:0934-9723
1435-4373
DOI:10.1007/BF01972003