Clinical evaluation of lysis-centrifugation technique and a biphasic bottle system for blood culture
The lysis centrifugation technique (Isolator, DuPont) for blood culture was compared with a system with biphasic medium in bottles. The Isolator was filled with 10 ml of blood once. One aerobic and one anaerobic bottle were injected 3 times with 2.5 ml of blood each. Organisms were detected in 90/74...
Gespeichert in:
Veröffentlicht in: | Scandinavian journal of infectious diseases 1985, Vol.17 (4), p.401-406 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The lysis centrifugation technique (Isolator, DuPont) for blood culture was compared with a system with biphasic medium in bottles. The Isolator was filled with 10 ml of blood once. One aerobic and one anaerobic bottle were injected 3 times with 2.5 ml of blood each. Organisms were detected in 90/748 blood cultures; 26 of which were contaminants. 34 pathogens were detected by both methods, 12 with the Isolator only and 18 with the 3 bottle pairs only. The first pair of bottles revealed 45/52 isolates, the second and third pairs gave an additional 6 and 1 isolate respectively. The contamination rate was 2.3% for the Isolator, which is lower than earlier reported, and 1.3% for the bottles. The most common pathogens were Escherichia coli, Pseudomonas aeruginosa and Streptococcus pneumoniae. The Isolator gave a faster diagnosis in 23 of the 34 cases. No decrease in the recovery rate was seen after 8 h, the longest recommended transport time for the Isolator tubes. One Isolator gave the same yield as the first pair of bottles and combining the methods increased the yield 25% compared to either method alone. |
---|---|
ISSN: | 0036-5548 1651-1980 |