Evaluation of the new DrySlide beta-lactamase test
Four hundred thirteen recent clinical isolates of Staphylococcus, Haemophilus, Moraxella, and Neisseria species were tested for beta-lactamase production using the Difco DrySlide beta-lactamase test. These results were compared with those of the BBL Cefinase reagent. Of the 413 isolates tested, 258...
Gespeichert in:
Veröffentlicht in: | American journal of clinical pathology 1994-06, Vol.101 (6), p.726-728 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Four hundred thirteen recent clinical isolates of Staphylococcus, Haemophilus, Moraxella, and Neisseria species were tested for beta-lactamase production using the Difco DrySlide beta-lactamase test. These results were compared with those of the BBL Cefinase reagent. Of the 413 isolates tested, 258 (62.5%) were beta-lactamase positive. There was 99.8% agreement between the two test methods; only one isolate of Staphylococcus aureus was DrySlide-negative and Cefinase-positive. The DrySlide beta-lactamase reagent is an accurate and convenient format for the evaluation of beta-lactamase activity in the clinical laboratory. |
---|---|
ISSN: | 0002-9173 1943-7722 |
DOI: | 10.1093/ajcp/101.6.726 |