A High-Performance Liquid Chromatography Method for Measurement of Oxidized Glutathione in Biological Samples
A high-performance liquid chromatography method to determine oxidized glutathione (GSSG) in biological samples with ultraviolet-visible detection using N-ethylmaleimide to prevent reduced glutathione (GSH) oxidation is described. Previous methods based on high-performance liquid chromatography to qu...
Gespeichert in:
Veröffentlicht in: | Analytical biochemistry 1994-03, Vol.217 (2), p.323-328 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A high-performance liquid chromatography method to determine oxidized glutathione (GSSG) in biological samples with ultraviolet-visible detection using N-ethylmaleimide to prevent reduced glutathione (GSH) oxidation is described. Previous methods based on high-performance liquid chromatography to quantitate GSH and GSSG are unsuitable for determining GSSG in biological samples. This is due to GSH oxidation during sample processing. N-Ethylmaleimide, but not iodacetic acid, prevents this oxidation. Blood GSH oxidation measured by the widely used method of Reed et al. (Anal. Biochem. 106, 55-62, 1980) can be as high as 24 ± 6% (n = 6). When blood samples were assayed by our procedure, GSH oxidation was only 0.13 ± 0.28% (n = 5). GSH can be determined enzymatically, i.e., with glutathione-S-transferase, but perchloric acid should not be used to deproteinize samples. Trichloroacetic acid (15% final concentration) may be used. This method allows an accurate calculation of the GSH/GSSG ratio, which is important for determining oxidative stress in tissues in various pathophysiological situations. |
---|---|
ISSN: | 0003-2697 1096-0309 |
DOI: | 10.1006/abio.1994.1126 |