Effect of Transvenous Atrioventricular Nodal Ablation on the Function of Implanted Pacemakers

We report on seven patients with implanted pacemakers who underwent transvenous ablation of the atrioventricular junction using direct current shocks of 200 to 350 joules. Pacemaker impulse duration and rate were unaffected, but one rate responsive (TX) pacemaker was reprogrammed by a 300‐joule shoc...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 1985-11, Vol.8 (6), p.811-814
Hauptverfasser: BOWES, ROBERT J., BENNETT, DAVID H.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report on seven patients with implanted pacemakers who underwent transvenous ablation of the atrioventricular junction using direct current shocks of 200 to 350 joules. Pacemaker impulse duration and rate were unaffected, but one rate responsive (TX) pacemaker was reprogrammed by a 300‐joule shock. Transient increases in stimulation threshold did occur in two patients, and exit block for 2–15 seconds developed on four occasions. Chronic stimulation thresholds were unaffected. We conclude that it is preferable to carry out ablation before pacemaker implantation, but it is possible to perform transvenous ablation of the atrioventricular junction without damage to an implanted pacemaker; however, a transient rise in stimulation threshold or even exit block may occur, and pacemaker function should be carefully assessed after the procedure.
ISSN:0147-8389
1540-8159
DOI:10.1111/j.1540-8159.1985.tb05899.x