Unrecognized hazards of surgical electrodes passed through metal suction-irrigation devices
Surgical electrodes, passed through metal suction-irrigation devices, pose significant risks for unrecognized visceral burns through capacitively coupled current using monopolar electrosurgery. Plastic cannulas (and reducers) should be avoided with the metal suction-irrigation electrode; an all-meta...
Gespeichert in:
Veröffentlicht in: | Surgical endoscopy 1994-03, Vol.8 (3), p.185-187 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Surgical electrodes, passed through metal suction-irrigation devices, pose significant risks for unrecognized visceral burns through capacitively coupled current using monopolar electrosurgery. Plastic cannulas (and reducers) should be avoided with the metal suction-irrigation electrode; an all-metal trocar cannula confers limited safety. With surgeon education and advances in engineering, the potential for unrecognized visceral injury with capacitive coupling can be eliminated. |
---|---|
ISSN: | 0930-2794 1432-2218 |
DOI: | 10.1007/BF00591827 |