Use of pulse oximetry to monitor venous saturation during extracorporeal life support
OBJECTIVETo assess the ability of two different pulse oximeters to display continuous venous oxygen saturation through an extracorporeal bypass circuit with a degree of accuracy comparable to direct in-line oximetry. DESIGNProspective, comparison study of pulse oximeters (test oximeter 1 or test oxi...
Gespeichert in:
Veröffentlicht in: | Critical care medicine 1994-04, Vol.22 (4), p.626-632 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | OBJECTIVETo assess the ability of two different pulse oximeters to display continuous venous oxygen saturation through an extracorporeal bypass circuit with a degree of accuracy comparable to direct in-line oximetry.
DESIGNProspective, comparison study of pulse oximeters (test oximeter 1 or test oximeter 2) and an in-line oximeter (test oximeter 3).
SETTINGA tertiary care neonatal intensive care unit.
PATIENTSSixty-five consecutive neonates with severe cardiorespiratory failure undergoing extracorporeal life support.
INTERVENTIONSThe accuracy of the oximeters was determined by simultaneously comparing the saturation displayed by the pulse oximeters (test oximeters 1 and 2) and/or the in-line oximeter (test oximeter 3) with the measured fractional venous oxygen saturation obtained at regular intervals from the extracorporeal circuit.
MAIN OUTCOME MEASURESVenous oxygen saturation was the criterion standard used to determine accuracy. Bias was defined as the mean difference between observed pulse oximeter or in-line oximeter values and the measured venous oxygen saturation. Mean biases were calculated for venous oxygen saturation measurements between 55% and 99% at intervals of 10%. Precision (the standard deviation of the bias) was calculated for low (55% to 75%), medium (76% to 81%), and high (82% to 99%) venous oxygen saturation values. A total of 983 venous oxygen saturation measurements were made and compared with simultaneous oximeter readings from test oximeter 1 (n = 600), test oximeter 2 (n = 478), and test oximeter 3 (n = 587).
RESULTSTest oximeter 1 was the most precise instrument at each level of venous oxygen saturation (SD, 4.0 to 4.8). Test oximeter 3 demonstrated the most consistent mean bias (range, 8), but was the most inaccurate oximeter across all levels of venous oxygen saturation.
CONCLUSIONSIn addition to its known clinical usefulness, pulse oximetry may serve as an adequate substitute for in-line oximetry during extracorporeal life support. (Crit Care Med 1994; 22:626–632) |
---|---|
ISSN: | 0090-3493 1530-0293 |
DOI: | 10.1097/00003246-199404000-00019 |