Emissivity measurements of reflective surfaces at near-millimeter wavelengths

We have developed an instrument for directly measuring the emissivity of reflective surfaces at near-millimeter wavelengths. The thermal emission of a test sample is compared with that of a reference surface, allowing the emissivity of the sample to be determined without heating. The emissivity of t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied Optics 1995-08, Vol.34 (22), p.4812-4816
Hauptverfasser: Bock, J J, Parikh, M K, Fischer, M L, Lange, A E
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We have developed an instrument for directly measuring the emissivity of reflective surfaces at near-millimeter wavelengths. The thermal emission of a test sample is compared with that of a reference surface, allowing the emissivity of the sample to be determined without heating. The emissivity of the reference surface is determined by one's heating the reference surface and measuring the increase in emission. The instrument has an absolute accuracy of Δε = 5 × 10(-4) and can reproducibly measure a difference in emissivity as small as Δε = 10(-4) between flat reflective samples. We have used the instrument to measure the emissivity of metal films evaporated on glass and carbon fiber-reinforced plastic composite surfaces. We measure an emissivity of (2.15 ± 0.4) × 10(-3) for gold evaporated on glass and (2.65 ± 0.5) × 10(-3) for aluminum evaporated on carbon fiber-reinforced plastic composite.
ISSN:1559-128X
0003-6935
1539-4522
DOI:10.1364/AO.34.004812