A quality-control phantom is described which is capable of testing the performance of inhalation rebreathing systems for measurements of regional cerebral blood flow (rCBF). The phantom is designed so that data similar to those from a patient study are obtained. The standard rCBF data-processing alg...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of nuclear medicine (1978) 1985-07, Vol.26 (7), p.783-789
Hauptverfasser: Dunn, William L, Oswald, William M, Wahner, Heinz W
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A quality-control phantom is described which is capable of testing the performance of inhalation rebreathing systems for measurements of regional cerebral blood flow (rCBF). The phantom is designed so that data similar to those from a patient study are obtained. The standard rCBF data-processing algorithm is used for the calculation of phantom study results. Malfunctions were induced to simulate instrument malfunction, and their effects on detector and air curves were evaluated. The use of this phantom is recommended prior to a patient study if the rCBF instrument is used infrequently or after service for routine maintenance or malfunction.