In vitro evaluation of a continuous-sampling device for pharmacokinetic parameter estimation

Using an in vitro pharmacokinetic model, area under the curve (AUC) estimates from an osmotic continuous-withdrawal device were compared to AUC estimates by a conventional trapezoidal method. Ten experiments were done under two different conditions: (a) half-life of 1 h (n = 5) and (b) half-life of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Therapeutic drug monitoring 1993-08, Vol.15 (4), p.294-299
Hauptverfasser: PAAP, C. M, CHOCAS, E. C
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 299
container_issue 4
container_start_page 294
container_title Therapeutic drug monitoring
container_volume 15
creator PAAP, C. M
CHOCAS, E. C
description Using an in vitro pharmacokinetic model, area under the curve (AUC) estimates from an osmotic continuous-withdrawal device were compared to AUC estimates by a conventional trapezoidal method. Ten experiments were done under two different conditions: (a) half-life of 1 h (n = 5) and (b) half-life of 2 h (n = 5). Sampling was done for 6 and 12 h, respectively. The AUC estimates from the two methods were highly correlated (r = 0.948, p < 0.0001). The mean coefficient of variation was 9.5% (n = 10), and mean AUC values were not statistically different by analysis of variance (p = 0.4). The sampling device sampled at 0.121 +/- 0.011 ml/h; the mean volume was 0.79 +/- 0.03 and 1.34 +/- 0.02 ml over 6 and 12 h, respectively. Continuous sampling provided a reasonable estimation of AUC, required only one analytical sample, and sampled at a consistent zero-order rate.
doi_str_mv 10.1097/00007691-199308000-00006
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_76082451</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>76082451</sourcerecordid><originalsourceid>FETCH-LOGICAL-c284t-2dbc6bc3e0b1c90117ca41af02a376857f6368b9fa799b0fa7cc315c5c51c1ab3</originalsourceid><addsrcrecordid>eNo9UMFO5DAMjVYgGNj9BKQcELdC3KRpckQIdpGQuMBtpcr1pBBom5K0I-3fbwaGsQ-W7ffsp8cYB3EJwtZXIketLRRgrRQmd8V2pH-wFVRSF1JbdcBWArQslKz0MTtJ6U0IUBl7xI5MKbXUasX-3o984-cYuNtgv-Dsw8hDx5FTGGc_LmFJRcJh6v34wtdu48nxLkQ-vWIckMK7H93siU8YcXCzi9yl2Q-fh36yww775H7t6il7vrt9uvlTPDz-vr-5fiioNGouynVLuiXpRAtkBUBNqAA7UaKstanqLms1re2wtrYVuRBJqCgnEGArT9nF190pho8l_28Gn8j1PY4u629qLUypKshA8wWkGFKKrmummLXGfw2IZmts821sszf2c6Qz9Wz3Y2kHt94Td07m_fluj4mw7yKO5NMepoxSUBn5H89qgfo</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>76082451</pqid></control><display><type>article</type><title>In vitro evaluation of a continuous-sampling device for pharmacokinetic parameter estimation</title><source>MEDLINE</source><source>Journals@Ovid Complete</source><creator>PAAP, C. M ; CHOCAS, E. C</creator><creatorcontrib>PAAP, C. M ; CHOCAS, E. C</creatorcontrib><description>Using an in vitro pharmacokinetic model, area under the curve (AUC) estimates from an osmotic continuous-withdrawal device were compared to AUC estimates by a conventional trapezoidal method. Ten experiments were done under two different conditions: (a) half-life of 1 h (n = 5) and (b) half-life of 2 h (n = 5). Sampling was done for 6 and 12 h, respectively. The AUC estimates from the two methods were highly correlated (r = 0.948, p &lt; 0.0001). The mean coefficient of variation was 9.5% (n = 10), and mean AUC values were not statistically different by analysis of variance (p = 0.4). The sampling device sampled at 0.121 +/- 0.011 ml/h; the mean volume was 0.79 +/- 0.03 and 1.34 +/- 0.02 ml over 6 and 12 h, respectively. Continuous sampling provided a reasonable estimation of AUC, required only one analytical sample, and sampled at a consistent zero-order rate.</description><identifier>ISSN: 0163-4356</identifier><identifier>EISSN: 1536-3694</identifier><identifier>DOI: 10.1097/00007691-199308000-00006</identifier><identifier>PMID: 8236364</identifier><identifier>CODEN: TDMODV</identifier><language>eng</language><publisher>Hagerstown, MD: Lippincott Williams &amp; Wilkins</publisher><subject>Biological and medical sciences ; Drug Monitoring - instrumentation ; Evaluation Studies as Topic ; General pharmacology ; Half-Life ; Humans ; Medical sciences ; Methylene Blue - analysis ; Models, Biological ; Pharmacokinetics ; Pharmacokinetics. Pharmacogenetics. Drug-receptor interactions ; Pharmacology. Drug treatments ; Regression Analysis ; Software ; Specimen Handling - instrumentation</subject><ispartof>Therapeutic drug monitoring, 1993-08, Vol.15 (4), p.294-299</ispartof><rights>1993 INIST-CNRS</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27923,27924</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&amp;idt=4844158$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/8236364$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>PAAP, C. M</creatorcontrib><creatorcontrib>CHOCAS, E. C</creatorcontrib><title>In vitro evaluation of a continuous-sampling device for pharmacokinetic parameter estimation</title><title>Therapeutic drug monitoring</title><addtitle>Ther Drug Monit</addtitle><description>Using an in vitro pharmacokinetic model, area under the curve (AUC) estimates from an osmotic continuous-withdrawal device were compared to AUC estimates by a conventional trapezoidal method. Ten experiments were done under two different conditions: (a) half-life of 1 h (n = 5) and (b) half-life of 2 h (n = 5). Sampling was done for 6 and 12 h, respectively. The AUC estimates from the two methods were highly correlated (r = 0.948, p &lt; 0.0001). The mean coefficient of variation was 9.5% (n = 10), and mean AUC values were not statistically different by analysis of variance (p = 0.4). The sampling device sampled at 0.121 +/- 0.011 ml/h; the mean volume was 0.79 +/- 0.03 and 1.34 +/- 0.02 ml over 6 and 12 h, respectively. Continuous sampling provided a reasonable estimation of AUC, required only one analytical sample, and sampled at a consistent zero-order rate.</description><subject>Biological and medical sciences</subject><subject>Drug Monitoring - instrumentation</subject><subject>Evaluation Studies as Topic</subject><subject>General pharmacology</subject><subject>Half-Life</subject><subject>Humans</subject><subject>Medical sciences</subject><subject>Methylene Blue - analysis</subject><subject>Models, Biological</subject><subject>Pharmacokinetics</subject><subject>Pharmacokinetics. Pharmacogenetics. Drug-receptor interactions</subject><subject>Pharmacology. Drug treatments</subject><subject>Regression Analysis</subject><subject>Software</subject><subject>Specimen Handling - instrumentation</subject><issn>0163-4356</issn><issn>1536-3694</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1993</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNo9UMFO5DAMjVYgGNj9BKQcELdC3KRpckQIdpGQuMBtpcr1pBBom5K0I-3fbwaGsQ-W7ffsp8cYB3EJwtZXIketLRRgrRQmd8V2pH-wFVRSF1JbdcBWArQslKz0MTtJ6U0IUBl7xI5MKbXUasX-3o984-cYuNtgv-Dsw8hDx5FTGGc_LmFJRcJh6v34wtdu48nxLkQ-vWIckMK7H93siU8YcXCzi9yl2Q-fh36yww775H7t6il7vrt9uvlTPDz-vr-5fiioNGouynVLuiXpRAtkBUBNqAA7UaKstanqLms1re2wtrYVuRBJqCgnEGArT9nF190pho8l_28Gn8j1PY4u629qLUypKshA8wWkGFKKrmummLXGfw2IZmts821sszf2c6Qz9Wz3Y2kHt94Td07m_fluj4mw7yKO5NMepoxSUBn5H89qgfo</recordid><startdate>19930801</startdate><enddate>19930801</enddate><creator>PAAP, C. M</creator><creator>CHOCAS, E. C</creator><general>Lippincott Williams &amp; Wilkins</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>19930801</creationdate><title>In vitro evaluation of a continuous-sampling device for pharmacokinetic parameter estimation</title><author>PAAP, C. M ; CHOCAS, E. C</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c284t-2dbc6bc3e0b1c90117ca41af02a376857f6368b9fa799b0fa7cc315c5c51c1ab3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1993</creationdate><topic>Biological and medical sciences</topic><topic>Drug Monitoring - instrumentation</topic><topic>Evaluation Studies as Topic</topic><topic>General pharmacology</topic><topic>Half-Life</topic><topic>Humans</topic><topic>Medical sciences</topic><topic>Methylene Blue - analysis</topic><topic>Models, Biological</topic><topic>Pharmacokinetics</topic><topic>Pharmacokinetics. Pharmacogenetics. Drug-receptor interactions</topic><topic>Pharmacology. Drug treatments</topic><topic>Regression Analysis</topic><topic>Software</topic><topic>Specimen Handling - instrumentation</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>PAAP, C. M</creatorcontrib><creatorcontrib>CHOCAS, E. C</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Therapeutic drug monitoring</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>PAAP, C. M</au><au>CHOCAS, E. C</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>In vitro evaluation of a continuous-sampling device for pharmacokinetic parameter estimation</atitle><jtitle>Therapeutic drug monitoring</jtitle><addtitle>Ther Drug Monit</addtitle><date>1993-08-01</date><risdate>1993</risdate><volume>15</volume><issue>4</issue><spage>294</spage><epage>299</epage><pages>294-299</pages><issn>0163-4356</issn><eissn>1536-3694</eissn><coden>TDMODV</coden><abstract>Using an in vitro pharmacokinetic model, area under the curve (AUC) estimates from an osmotic continuous-withdrawal device were compared to AUC estimates by a conventional trapezoidal method. Ten experiments were done under two different conditions: (a) half-life of 1 h (n = 5) and (b) half-life of 2 h (n = 5). Sampling was done for 6 and 12 h, respectively. The AUC estimates from the two methods were highly correlated (r = 0.948, p &lt; 0.0001). The mean coefficient of variation was 9.5% (n = 10), and mean AUC values were not statistically different by analysis of variance (p = 0.4). The sampling device sampled at 0.121 +/- 0.011 ml/h; the mean volume was 0.79 +/- 0.03 and 1.34 +/- 0.02 ml over 6 and 12 h, respectively. Continuous sampling provided a reasonable estimation of AUC, required only one analytical sample, and sampled at a consistent zero-order rate.</abstract><cop>Hagerstown, MD</cop><pub>Lippincott Williams &amp; Wilkins</pub><pmid>8236364</pmid><doi>10.1097/00007691-199308000-00006</doi><tpages>6</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0163-4356
ispartof Therapeutic drug monitoring, 1993-08, Vol.15 (4), p.294-299
issn 0163-4356
1536-3694
language eng
recordid cdi_proquest_miscellaneous_76082451
source MEDLINE; Journals@Ovid Complete
subjects Biological and medical sciences
Drug Monitoring - instrumentation
Evaluation Studies as Topic
General pharmacology
Half-Life
Humans
Medical sciences
Methylene Blue - analysis
Models, Biological
Pharmacokinetics
Pharmacokinetics. Pharmacogenetics. Drug-receptor interactions
Pharmacology. Drug treatments
Regression Analysis
Software
Specimen Handling - instrumentation
title In vitro evaluation of a continuous-sampling device for pharmacokinetic parameter estimation
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-08T15%3A35%3A53IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=In%20vitro%20evaluation%20of%20a%20continuous-sampling%20device%20for%20pharmacokinetic%20parameter%20estimation&rft.jtitle=Therapeutic%20drug%20monitoring&rft.au=PAAP,%20C.%20M&rft.date=1993-08-01&rft.volume=15&rft.issue=4&rft.spage=294&rft.epage=299&rft.pages=294-299&rft.issn=0163-4356&rft.eissn=1536-3694&rft.coden=TDMODV&rft_id=info:doi/10.1097/00007691-199308000-00006&rft_dat=%3Cproquest_cross%3E76082451%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=76082451&rft_id=info:pmid/8236364&rfr_iscdi=true