In vitro evaluation of a continuous-sampling device for pharmacokinetic parameter estimation
Using an in vitro pharmacokinetic model, area under the curve (AUC) estimates from an osmotic continuous-withdrawal device were compared to AUC estimates by a conventional trapezoidal method. Ten experiments were done under two different conditions: (a) half-life of 1 h (n = 5) and (b) half-life of...
Gespeichert in:
Veröffentlicht in: | Therapeutic drug monitoring 1993-08, Vol.15 (4), p.294-299 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Using an in vitro pharmacokinetic model, area under the curve (AUC) estimates from an osmotic continuous-withdrawal device were compared to AUC estimates by a conventional trapezoidal method. Ten experiments were done under two different conditions: (a) half-life of 1 h (n = 5) and (b) half-life of 2 h (n = 5). Sampling was done for 6 and 12 h, respectively. The AUC estimates from the two methods were highly correlated (r = 0.948, p < 0.0001). The mean coefficient of variation was 9.5% (n = 10), and mean AUC values were not statistically different by analysis of variance (p = 0.4). The sampling device sampled at 0.121 +/- 0.011 ml/h; the mean volume was 0.79 +/- 0.03 and 1.34 +/- 0.02 ml over 6 and 12 h, respectively. Continuous sampling provided a reasonable estimation of AUC, required only one analytical sample, and sampled at a consistent zero-order rate. |
---|---|
ISSN: | 0163-4356 1536-3694 |
DOI: | 10.1097/00007691-199308000-00006 |