In vitro evaluation of a continuous-sampling device for pharmacokinetic parameter estimation

Using an in vitro pharmacokinetic model, area under the curve (AUC) estimates from an osmotic continuous-withdrawal device were compared to AUC estimates by a conventional trapezoidal method. Ten experiments were done under two different conditions: (a) half-life of 1 h (n = 5) and (b) half-life of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Therapeutic drug monitoring 1993-08, Vol.15 (4), p.294-299
Hauptverfasser: PAAP, C. M, CHOCAS, E. C
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Using an in vitro pharmacokinetic model, area under the curve (AUC) estimates from an osmotic continuous-withdrawal device were compared to AUC estimates by a conventional trapezoidal method. Ten experiments were done under two different conditions: (a) half-life of 1 h (n = 5) and (b) half-life of 2 h (n = 5). Sampling was done for 6 and 12 h, respectively. The AUC estimates from the two methods were highly correlated (r = 0.948, p < 0.0001). The mean coefficient of variation was 9.5% (n = 10), and mean AUC values were not statistically different by analysis of variance (p = 0.4). The sampling device sampled at 0.121 +/- 0.011 ml/h; the mean volume was 0.79 +/- 0.03 and 1.34 +/- 0.02 ml over 6 and 12 h, respectively. Continuous sampling provided a reasonable estimation of AUC, required only one analytical sample, and sampled at a consistent zero-order rate.
ISSN:0163-4356
1536-3694
DOI:10.1097/00007691-199308000-00006