Biomechanical Stress Distribution on Fixation Screws Used in Bilateral Sagittal Split Ramus Osteotomy: Assessment of 9 Methods via Finite Element Method

Purpose The aim of this study was to assess the biomechanical stress tolerance of screws used in 9 fixation methods after bilateral sagittal split ramus osteotomy to determine which configuration leads to lesser force load on the cortical bone at fixation points. Materials and Methods A 3-dimensiona...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of oral and maxillofacial surgery 2010-11, Vol.68 (11), p.2765-2769
Hauptverfasser: Bohluli, Behnam, DMD, Motamedi, Mohammad Hosein Kalantar, DDS, Bohluli, Pedram, DDS, MS, PhD, Sarkarat, Farzin, DMD, Moharamnejad, Nima, DMD, Tabrizi, Mohammad Hossein Seif, DDS
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose The aim of this study was to assess the biomechanical stress tolerance of screws used in 9 fixation methods after bilateral sagittal split ramus osteotomy to determine which configuration leads to lesser force load on the cortical bone at fixation points. Materials and Methods A 3-dimensional computerized model of a human mandible with posterior teeth was generated. The bilateral sagittal split ramus osteotomy was virtually performed on this model. The separated model was assembled with 9 fixation methods: single screw, 2 screws one behind the other, 2 screws one below the other, 3 screws in an L configuration, 3 screws in an inverted backward L configuration, miniplate with 2 screws, miniplate with 4 screws, 2 parallel plates (upper + lower border), and square miniplate with 4 screws. Then, 75-, 135-, and 600-N vertical loads were applied on the posterior teeth of these models. The stress distribution on the screw sites on the buccal cortex was measured by the finite element method. Results In this model all the fixation methods withstood forces between 75 and 135 N. However, the single-screw and the 2-hole miniplate models showed that the stress distributions in the configurations were intolerable when 600 N of posterior force was applied. The results of this study indicated that the inverted backward L configuration with 3 bicortical screws was the most stable. Conclusion Although this study indicated that the inverted backward L configuration with 3 bicortical screws was the most stable pattern, most of the patterns had adequate stability for clinical applications (mean, 125 N).