Inter-comparison of 18F-FDG PET/CT standardized uptake values in Korea
The number of PET/CT tests for diagnosing and treating tumors are increasing every year, and thus it is considered necessary to compare PET/CT images among medical institutions or among different PET/CT systems within the same medical institution. We present to get the variation of standardized upta...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2011, Vol.69 (1), p.241-246 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The number of PET/CT tests for diagnosing and treating tumors are increasing every year, and thus it is considered necessary to compare PET/CT images among medical institutions or among different PET/CT systems within the same medical institution. We present to get the variation of standardized uptake values (SUV) among PET/CT systems, to derive an SUV calibration table that enables quantitative comparison of PET/CT systems, and to test the usefulness of the calibration table through comparing patient’s SUV with the SUV calibration table. The correlation coefficients that inter-compared the calibration constants of normal patients and the calibration constants of the mean standardized uptake value with the calibration constants of the standardized uptake value of phantom are 0.88938 (