Comparison of skin necrosis in rats by using a new microneedle electrocautery, standard-size needle electrocautery, and the shaw hemostatic scalpel

"Microneedle" electrocautery was compared against the standard-size needle electrocautery and the Shaw hemostatic scalpel to determine the differences in tissue necrosis when used as a cutting instrument. Incisions were made on the dorsal skin of anesthetized white rats using each of the t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annals of plastic surgery 1993-08, Vol.31 (2), p.164-167
Hauptverfasser: FARNWORTH, T. K, BEALS, S. P, MANWARING, K. H, TREPETA, R. W
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:"Microneedle" electrocautery was compared against the standard-size needle electrocautery and the Shaw hemostatic scalpel to determine the differences in tissue necrosis when used as a cutting instrument. Incisions were made on the dorsal skin of anesthetized white rats using each of the three devices with the no. 15 scalpel as control. The specimens were submitted for histological evaluation. The microneedle caused less necrosis than the standard-size needle electrocautery (0.18 vs 0.27 mm, p < 0.01) and less necrosis than the Shaw hemostatic scalpel set at 220 degrees F (0.18 vs 0.25 mm, p < 0.05). The microneedle electrocautery was also found to be an instrument that causes very little tissue distortion during fine dissection and helps to minimize blood loss in craniofacial and neurosurgical operations.
ISSN:0148-7043
1536-3708
DOI:10.1097/00000637-199308000-00016