Guiding Catheter Tip Avulsion—An Unusual Complication of Renal Artery Stenting

A 68-year-old hypertensive male underwent stenting for critical stenosis of the right renal artery through right femoral approach. The stent missed the ostium and repeated attempts at stenting the ostium did not succeed. Brachial approach was resorted to and ostium could be stented successfully. Whi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Heart, lung & circulation lung & circulation, 2010-11, Vol.19 (11), p.675-677
Hauptverfasser: Subban, Vijayakumar, DM, George, Thomas, DNB, Sankardas, Mullasari Ajit, DM, Janakiraman, Ezhilan, DM
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A 68-year-old hypertensive male underwent stenting for critical stenosis of the right renal artery through right femoral approach. The stent missed the ostium and repeated attempts at stenting the ostium did not succeed. Brachial approach was resorted to and ostium could be stented successfully. While deploying the stent the balloon was slightly inside the guiding catheter and the catheter tip got partially detached. The whole assembly was withdrawn to the brachial artery and the detached fragment was snared and removed.
ISSN:1443-9506
1444-2892
DOI:10.1016/j.hlc.2010.06.1055