CT of a Bronchial Phantom: Factors Affecting Appearance and Size Measurements

In order to determine the most appropriate window settings for viewing CT of the bronchial tree, we performed CT of a bronchial phantom consisting of air-filled tubes measuring from 3.1 to 12.7 mm, oriented at varying angles relative to the scan plane, surrounded by water or air, and with scan colli...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Investigative radiology 1984-09, Vol.19 (5), p.394-398
Hauptverfasser: WEBB, W RICHARD, GAMSU, GORDON, WALL, SUSAN D, CANN, CHRISTOPHER E, PROCTOR, EVELYN
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In order to determine the most appropriate window settings for viewing CT of the bronchial tree, we performed CT of a bronchial phantom consisting of air-filled tubes measuring from 3.1 to 12.7 mm, oriented at varying angles relative to the scan plane, surrounded by water or air, and with scan collimation of 10 mm, 5 mm, and 1.5 mm. Using a computer program to graphically display CT number relative to the distance across the tubeʼs lumen, it was found that a window mean of −150 H accurately estimated the internal diameter of tubes surrounded by water, at all angles, when collimation was 5 mm or 1.5 mm. With 10-mm collimation, tube diameter was slightly underestimated for tubes 9.5 mm or less when oriented 30° or more from perpendicular to the plane of scan. At lower window settings and window widths of 500 H or less, all tubeʼs diameters were significantly underestimated. At –150 H, with tubes parallel to and centered in the scan plane, 5-mm and 1.5-mm collimation were most accurate; with decenter-ing of 4 mm, 10-mm collimation better showed the tubeʼs lumen. When surrounded by air, tube wall thickness was best estimated using a window mean of −450 H.
ISSN:0020-9996
1536-0210
DOI:10.1097/00004424-198409000-00011