Iterative Reconstruction Methods for Nonuniform Attenuation Distribution in SPECT

Two iterative methods, a generalization of the Chang method and a projection precorrection, were investigated to determine whether the use of an attenuation map could improve nonuniform attenuation compensation. After a detailed description of the methods, results obtained with simulated and phantom...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of nuclear medicine (1978) 1993-07, Vol.34 (7), p.1204-1209
Hauptverfasser: Maze, Anne, Le Cloirec, Joseph, Collorec, Rene, Bizais, Yves, Briandet, Philippe, Bourguet, Patrick
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Two iterative methods, a generalization of the Chang method and a projection precorrection, were investigated to determine whether the use of an attenuation map could improve nonuniform attenuation compensation. After a detailed description of the methods, results obtained with simulated and phantom data were compared. This study demonstrates that projection precorrection provides accurate quantification and good image quality as early as the precorrection step, whereas the generalized Chang method requires computation of one more iteration.
ISSN:0161-5505
1535-5667