Is There an Effect of Amiodarone on the Defibrillation Threshold?

The interaction between amiodarone and the defibrillation threshold (DFT) is still controversial. We present a case with dilated cardiomyopathy and recurrent sustained monomorphic ventricular tachycardia who received an automatic implantable cardioverter defibrillator (AICD) while under long-term am...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Japanese Heart Journal 1993, Vol.34(2), pp.221-226
Hauptverfasser: GÜLDAL, Muharrem, KARAOGUZ, Remzi, AKALIN, Hakki, BAYAR, Mustafa, AKYOL, Turhan
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The interaction between amiodarone and the defibrillation threshold (DFT) is still controversial. We present a case with dilated cardiomyopathy and recurrent sustained monomorphic ventricular tachycardia who received an automatic implantable cardioverter defibrillator (AICD) while under long-term amiodarone treatment. AICD implantation was performed without thoracotomy. The transvenous lead was inserted via a left subclavian vein puncture and the patch was placed on the lateral chest wall, submuscularly. At the time of implantation a 35J shock was not successful in converting ventricular fibrillation to sinus rhythm, but a 40J rescue shock was successful. After discontinuation of amiodarone, DFT measurements were repeated. Sixteen days later DFT was still higher than 34J, but 71 days later it decreased to 20J.
ISSN:0021-4868
1348-673X
DOI:10.1536/ihj.34.221