Effects of Characteristic X-Rays on Assay of I-123 by Dose Calibrator
Large differences in dose-calibrator readings are obtained if "high-purity" I-123 is assayed in different containers. Large correction factors are necessary for assaying another isotope of iodine, I-125, in a dose calibrator, because of absorption of the low-energy (28.4-keV, weighted mean...
Gespeichert in:
Veröffentlicht in: | The Journal of nuclear medicine (1978) 1984-12, Vol.25 (12), p.1367-1370 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Large differences in dose-calibrator readings are obtained if "high-purity" I-123 is assayed in different containers. Large correction factors are necessary for assaying another isotope of iodine, I-125, in a dose calibrator, because of absorption of the low-energy (28.4-keV, weighted mean) emissions. We found that up to 70% of the dose-calibrator response to I-123 can be due to characteristic x-rays with energies exactly the same as those emitted by I-125, and that dose-calibrator response to I-123 is also strongly affected by the absorption properties of the vial. An appropriate method to define I-123 activity uses a gamma camera with a medium-energy collimator to establish correction factors for dose-calibrator assay of I-123 in different containers. Correction factors for a plastic syringe and a thick-wall glass vial, were determined using this method. Measurement of I-123 activity in a copper absorber will eliminate the response to x-rays, and the gamma camera is useful in establishing the necessary correction factors. |
---|---|