RESISTANCE TO AIRFLOW IN ANAESTHETIC BREATHING SYSTEMS
We have examined, under reproducible conditions, the resistance to airflow of complete anaesthetic breathing systems (Magill, Coaxial Lack, Parallel Lack and Bain systems) and components of these systems. All systems had resistances within the recommended ranges at all flows likely to be experienced...
Gespeichert in:
Veröffentlicht in: | British journal of anaesthesia : BJA 1989-04, Vol.62 (4), p.456-461 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We have examined, under reproducible conditions, the resistance to airflow of complete anaesthetic breathing systems (Magill, Coaxial Lack, Parallel Lack and Bain systems) and components of these systems. All systems had resistances within the recommended ranges at all flows likely to be experienced in normal clinical practice. The Magill system had the lowest resistance under all conditions. It is concluded that comparisons should be made only between complete breathing systems. |
---|---|
ISSN: | 0007-0912 1471-6771 |