Development of a long-lasting ventricular catheter impregnated with a combination of antibiotics

A ventricular silicone catheter impregnated with a combination of rifampin and a quinolone was developed in order to prevent ventricular shunt related infections. As model substance for the quinolones we used sparfloxacin, because of its specific physicochemical properties resulting in a quantitativ...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Biomaterials 2003-11, Vol.24 (26), p.4865-4869
Hauptverfasser: Kohnen, Wolfgang, Kolbenschlag, Christian, Teske-Keiser, Susanne, Jansen, Bernd
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A ventricular silicone catheter impregnated with a combination of rifampin and a quinolone was developed in order to prevent ventricular shunt related infections. As model substance for the quinolones we used sparfloxacin, because of its specific physicochemical properties resulting in a quantitative detection also in the presence of a second antibiotic. In our study we focused especially on an optimization of the antibiotic release out of the impregnated catheters in order to develop long lasting devices with a broad antimicrobial spectrum. A release-optimized catheter was tested with an in vitro colonization test and additionally with a method developed to examine the spread of bacteria on a catheter surface. In vitro experiments showed that the impregnated catheters reduce the colonization with Staphylococcus epidermidis for at least 1 year and prevent the spread of bacteria along the catheter surface.
ISSN:0142-9612
1878-5905
DOI:10.1016/S0142-9612(03)00379-X