The measurement of right ventricular ejection fraction by thermodilution : a comparison of values obtained using differing injectate ports
Thermodilution measurements of cardiac output and right ventricular ejection fraction were obtained using a rapid response pulmonary artery catheter. Values were compared when injectate was administered via either a cannula within the right internal jugular vein or the dedicated right atrial port of...
Gespeichert in:
Veröffentlicht in: | Anaesthesia 1993-04, Vol.48 (4), p.312-314 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Thermodilution measurements of cardiac output and right ventricular ejection fraction were obtained using a rapid response pulmonary artery catheter. Values were compared when injectate was administered via either a cannula within the right internal jugular vein or the dedicated right atrial port of the pulmonary artery catheter. Mean (SD) bias for cardiac output and right ventricular ejection fraction measurements were 0.08 (0.32) l.min-1 and 2.6 (6.6)% respectively. We therefore conclude that both injectate techniques will provide similar values for cardiac output but dissimilar values for right ventricular ejection fraction measurement. |
---|---|