Clinical and Hemodynamic Evaluation of the 19 mm Björk-Shiley Aortic Valve Prosthesis
Between November, 1973, and March, 1980, 43 patients underwent isolated aortic valve replacement with 19 mm Björk-Shiley prostheses at the Johns Hopkins Hospital. There were 4 male and 39 female patients ranging from 12 to 75 years old (mean, 54.5 years). Average weight was 62 ± 2 kg and average bod...
Gespeichert in:
Veröffentlicht in: | The Annals of thoracic surgery 1981-07, Vol.32 (1), p.50-57 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Between November, 1973, and March, 1980, 43 patients underwent isolated aortic valve replacement with 19 mm Björk-Shiley prostheses at the Johns Hopkins Hospital. There were 4 male and 39 female patients ranging from 12 to 75 years old (mean, 54.5 years). Average weight was 62 ± 2 kg and average body surface area, 1.64 ± 0.3 m
2. Five patients died within thirty days of operation; however, since 1975, hospital mortality has been 5.9%. The 38 survivors have been followed up for as long as 85 months (mean, 40 months). There were 4 late deaths, and actuarial survival in patients discharged from the hospital was 81% at five years. All long-term survivors were in New York Heart Association Functional Class I (29 patients) or Class II (5 patients). Preoperative and postoperative echocardiograms in 17 patients demonstrated significant decreases in mean left ventricular wall thickness (12.9 ± 1.8 mm vs 10.3 ± 1.4 mm;
p < 0.001) and in left ventricular mass (262 ± 95 gm vs 188 ± 50 gm;
p < 0.02).
Postoperative cardiac catheterization data were obtained from an additional 24 patients undergoing aortic valve replacement with the 19 mm Björk-Shiley prosthesis at the National Heart Institute. Average peak systolic gradient at rest was 16 mm Hg (range, 0 to 45 mm Hg) and was found to be directly related to body surface area (
r = 0.60,
p < 0.002). Average effective valve orifice area was 1.06 cm
2 (range, 0.63 to 2.02 cm
2).
For patients with small aortic roots, aortic valve replacement with the 19 mm Björk-Shiley valve is a satisfactory and, perhaps, preferable alternative to aortic annuloplasty to accommodate larger sized prostheses. |
---|---|