Head fixation device for reproducible position alignment in transmission CT and positron emission tomography

A head-positioning device allowing accurate position transference between computed tomography and positron emission tomography is described. A base plate is fixed to the patient's head through either an individual mold or screws into the calvarium. This base plate fits onto the headholders of t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of computer assisted tomography 1981-02, Vol.5 (1), p.136-141
Hauptverfasser: Bergström, M, Boëthius, J, Eriksson, L, Greitz, T, Ribbe, T, Widén, L
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A head-positioning device allowing accurate position transference between computed tomography and positron emission tomography is described. A base plate is fixed to the patient's head through either an individual mold or screws into the calvarium. This base plate fits onto the headholders of the two devices, thus ensuring that scans are taken at the same level. Using the regular software, the coordinates of an anatomic structure may be determined and transferred between the two images.
ISSN:0363-8715
DOI:10.1097/00004728-198102000-00027