Simple, sensitive and rapid HPLC–MS/MS method for the determination of cepharanthine in human plasma
A rapid, sensitive and specific method for the determination of cepharanthine in human plasma using high performance liquid chromatography coupled with tandem mass spectrometry (HPLC–MS/MS) was described. Cepharanthine and the internal standard (I.S.), telmisartan, were extracted from human plasma b...
Gespeichert in:
Veröffentlicht in: | Journal of chromatography. B, Analytical technologies in the biomedical and life sciences Analytical technologies in the biomedical and life sciences, 2010-10, Vol.878 (28), p.2923-2927 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A rapid, sensitive and specific method for the determination of cepharanthine in human plasma using high performance liquid chromatography coupled with tandem mass spectrometry (HPLC–MS/MS) was described. Cepharanthine and the internal standard (I.S.), telmisartan, were extracted from human plasma by methanol to precipitate the protein. A centrifuged upper layer was then evaporated and reconstituted with 100
μL methanol. Chromatographic separation was performed on an AGILENT XDB-C
8 column (150
mm
×
2.1
mm, 5.0
μm, Agilent, USA) using a gradient mobile phase with 1
mmol/L ammonium acetate in water with 0.05% formic acid and methanol. Detection and quantitation was performed by MS/MS using electrospray ionization (ESI) and multiple reaction monitoring (MRM) in the positive ion mode. The most intense [M+H]
+ MRM transition of cepharanthine at
m/
z 607.3
→
365.3 was used for quantitation and the transition at
m/
z 515.5
→
276.4 was used to monitor telmisartan. The calibration curve was linear within the concentration range of 0.5–200.0
ng/mL (
r
=
0.9994). The limit of quantification (LOQ) was 0.5
ng/mL. The extraction recovery was above 81.1%. The accuracy was higher than 92.3%. The intra- and inter-day precisions were less than 9.66%. The method was accurate, sensitive and simple and was successfully applied to a pharmacokinetic study after single intravenous administration of 50
mg cepharanthine in 12 healthy Chinese volunteers. |
---|---|
ISSN: | 1570-0232 1873-376X |
DOI: | 10.1016/j.jchromb.2010.08.026 |