The rebreathing characteristics of the Hudson oxygen mask

Summary A lung model was used to study the effect of varying minute ventilation, oxygen flow and inspiratory: expiratory ratio on the rebreathing chracteristics of the Hudson oxygen mask. The onset of rebreathing was detected using two different sets of criteria; those defined by Kain et al. and Hum...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Anaesthesia 1993-03, Vol.48 (3), p.239-242
Hauptverfasser: CAMPKIN, N.T.A., OOI, R.G., SONI, N.C.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Summary A lung model was used to study the effect of varying minute ventilation, oxygen flow and inspiratory: expiratory ratio on the rebreathing chracteristics of the Hudson oxygen mask. The onset of rebreathing was detected using two different sets of criteria; those defined by Kain et al. and Humphrey. The onset of rebreathing was also quantified in terms of the inspired volume of carbon dioxide and increases in deadspace. The magnitude of rebreathing was increased by reducing oxygen inflow and increasing minute ventilation. Variation in I:E ratio between 1:1 and 1:4 had little effect on rebreathing. Using the clinical criteria of Kain et al, the onset of rebreathing was detected when oxygen inflow/minute ventilation ratio was reduced below 0.78 on average. With the other criterion, rebreathing occurred when the ratio was 0.46.