Non-radiological technique for three-dimensional imaging of endoscopes
A novel system is described which images in three dimensions, the total configuration of a colonoscope without the use of conventional radiological techniques. A low intensity magnetic field is used in conjunction with a miniature inductive sensor. The system intrinsically safe and it is potentially...
Gespeichert in:
Veröffentlicht in: | The Lancet (British edition) 1993-03, Vol.341 (8847), p.719-722 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A novel system is described which images in three dimensions, the total configuration of a colonoscope without the use of conventional radiological techniques. A low intensity magnetic field is used in conjunction with a miniature inductive sensor. The system intrinsically safe and it is potentially inexpensive and capable of being used in a normal hospital environment. Clinical trials are described in which the system is validated in terms of its suitability forthe application. Magnetic and conventional X-ray images obtained ex vivo with the endoscope held in various configurations and comparisons in the patients confirm the practical applicability of the new system. |
---|---|
ISSN: | 0140-6736 1474-547X |
DOI: | 10.1016/0140-6736(93)90487-2 |