Tined Transvenous Endocardial Electrodes: Results of a Randomized Prospective Study

The early (30-day) dislodgment rate of standard-use flange-tipped and of tined endocardial electrodes was compared in a randomized prospective clinical trial. Four of 16 (25%) of the flange-tipped leads and none of the 18 tined leads dislodged within 30 days of implantation ( p = 0.01). We believe t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Annals of thoracic surgery 1981-03, Vol.31 (3), p.289-289
Hauptverfasser: Richardson, James V., Wright, Creighton B., Ehrenhaft, J.L.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The early (30-day) dislodgment rate of standard-use flange-tipped and of tined endocardial electrodes was compared in a randomized prospective clinical trial. Four of 16 (25%) of the flange-tipped leads and none of the 18 tined leads dislodged within 30 days of implantation ( p = 0.01). We believe that the tined electrode represents a major improvement in electrode design and is clearly superior to the flange-tipped electrode in reducing the incidence of early dislodgment.
ISSN:0003-4975
1552-6259
DOI:10.1016/S0003-4975(10)60948-4