Tined Transvenous Endocardial Electrodes: Results of a Randomized Prospective Study
The early (30-day) dislodgment rate of standard-use flange-tipped and of tined endocardial electrodes was compared in a randomized prospective clinical trial. Four of 16 (25%) of the flange-tipped leads and none of the 18 tined leads dislodged within 30 days of implantation ( p = 0.01). We believe t...
Gespeichert in:
Veröffentlicht in: | The Annals of thoracic surgery 1981-03, Vol.31 (3), p.289-289 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The early (30-day) dislodgment rate of standard-use flange-tipped and of tined endocardial electrodes was compared in a randomized prospective clinical trial. Four of 16 (25%) of the flange-tipped leads and none of the 18 tined leads dislodged within 30 days of implantation (
p = 0.01). We believe that the tined electrode represents a major improvement in electrode design and is clearly superior to the flange-tipped electrode in reducing the incidence of early dislodgment. |
---|---|
ISSN: | 0003-4975 1552-6259 |
DOI: | 10.1016/S0003-4975(10)60948-4 |