Evaluation of the Ves-Matic Cube 200 Erythrocyte Sedimentation Method: Comparison With Westergren-Based Methods

The erythrocyte sedimentation rate (ESR) is still a widely used parameter for acute phase inflammation. Recently, new methods based on direct undiluted measurement of ESR in a standard EDTA tube have been developed. We evaluated the analytic performance of one of these new methods, the Ves-Matic Cub...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of clinical pathology 2010-10, Vol.134 (4), p.653-660
Hauptverfasser: CURVERS, Joyce, KOOREN, Jurgen, LAAN, Maartje, VAN LIEROP, Edwin, VAN DE KERKHOF, Daan, SCHARNHORST, Volkher, HERRUER, Martien
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The erythrocyte sedimentation rate (ESR) is still a widely used parameter for acute phase inflammation. Recently, new methods based on direct undiluted measurement of ESR in a standard EDTA tube have been developed. We evaluated the analytic performance of one of these new methods, the Ves-Matic Cube 200 (Diesse Diagnostica Senese, Siena, Italy), and compared it with several established Westergren-based diluted methods. The Ves-Matic Cube 200 showed a poor correlation (r = 0.83) with the International Council for Standardization in Haematology Westergren reference method, mainly caused by a considerable negative bias at low ESR levels. Moreover, a random bias was found at higher ESR levels that correlated with hematocrit levels, suggesting a differential influence of packed cell volume on the Ves-Matic Cube 200 results compared with Westergren results. We conclude that the Ves-Matic Cube 200 method is not interchangeable with Westergren-based diluted methods and generates ESR results that are too deviant to be clinically acceptable.
ISSN:0002-9173
1943-7722
DOI:10.1309/AJCPMEWW62BGQHJH