Continuous cardiac output monitoring system
A continuous cardiac output monitoring system has been developed in the laboratory to allow the real-time measurement of the cardiac output. This form of continuous cardiac output measurement allows the doctor to view the beat-to-beat cardiac output and can be employed to measure artery constriction...
Gespeichert in:
Veröffentlicht in: | Medical & biological engineering & computing 2001, Vol.39 (1), p.101-104 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A continuous cardiac output monitoring system has been developed in the laboratory to allow the real-time measurement of the cardiac output. This form of continuous cardiac output measurement allows the doctor to view the beat-to-beat cardiac output and can be employed to measure artery constrictions as well. The sensor comprises a laser Doppler velocimeter and an impedance measurement unit. The laser Doppler velocimeter is capable of measuring bi-directional blood flow within the vessel while the impedance measurement unit determines the cross-sectional area of the vessel. In laboratory tests, it was demonstrated on a heart-lung machine that the product of the two parameters measured is proportional to the actual flow volume of up to 6 lmin(-1) with a mean percentage error of 12.4% and a mean square error of 0.09 (using the lmin(-1) scale) were obtained. This is significantly more accurate than the measurement made using the thermodilution cathether. |
---|---|
ISSN: | 0140-0118 1741-0444 |
DOI: | 10.1007/BF02345272 |