Structured sleeve for repair of implantable in-line connectors

A structured miniature repair sleeve has been designed for implantable in-line connectors that develop small current leaks post-implant. The repair sleeve has been successfully utilised in one subject following the development of current leakage in connectors on an implanted joint angle sensor.

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical & biological engineering & computing 2000-07, Vol.38 (4), p.473-475
Hauptverfasser: STROJNIK, P, STAGE, T. G, BHADRA, N, KILGORE, K. L
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A structured miniature repair sleeve has been designed for implantable in-line connectors that develop small current leaks post-implant. The repair sleeve has been successfully utilised in one subject following the development of current leakage in connectors on an implanted joint angle sensor.
ISSN:0140-0118
1741-0444
DOI:10.1007/BF02345019