Large-Head Metal-on-Metal Total Hip Arthroplasty Using the Durom Acetabular Component at Minimum 1-Year Interval
Abstract Large-head metal-on-metal total hip arthroplasty represents novel technology, and outcome data are lacking. We prospectively compared the performance of a nonmodular metal-on-metal acetabular component (Durom; Zimmer, Warsaw, IN) with a modular titanium component (Trilogy, Zimmer). All Duro...
Gespeichert in:
Veröffentlicht in: | The Journal of arthroplasty 2010-09, Vol.25 (6), p.26-30 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract Large-head metal-on-metal total hip arthroplasty represents novel technology, and outcome data are lacking. We prospectively compared the performance of a nonmodular metal-on-metal acetabular component (Durom; Zimmer, Warsaw, IN) with a modular titanium component (Trilogy, Zimmer). All Durom components placed at our institution with minimum 1-year follow up (n = 63) were compared with an age- and sex-matched Trilogy control group (n = 100). Failure defined as revision or persistent moderate/severe groin pain was significantly higher for the Durom (11.1%) compared with the Trilogy group (0%) ( P = .002). Although all acetabular components in both groups appeared radiographically stable, no significant bone ingrowth was noted at the time of Durom revisions. We could not identify any patient/surgical-related factors predictive of failure. Further study is needed to determine the scientific basis for these observations. |
---|---|
ISSN: | 0883-5403 1532-8406 |
DOI: | 10.1016/j.arth.2010.04.005 |