Determination of optimum measurement points via A-optimality criterion for the calibration of measurement apparatus

A procedure for optimal selection of the measurement points to get the best calibration characteristics (for the chosen optimality criterion) of measuring apparatus is proposed. The coefficients of the calibration characteristics are evaluated by the classical least squares method. For this work, th...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Measurement : journal of the International Measurement Confederation 2010-05, Vol.43 (4), p.563-569
1. Verfasser: Hajiyev, Ch
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A procedure for optimal selection of the measurement points to get the best calibration characteristics (for the chosen optimality criterion) of measuring apparatus is proposed. The coefficients of the calibration characteristics are evaluated by the classical least squares method. For this work, the A-optimality criterion has been used as an optimality criteria. As an example, the problem of optimal selection of the standard pressure setters (the piston gauges) during calibration of the differential pressure gage is solved. Obtained values of the optimum measurement points for the calibration of the differential pressure gage are checked via actual experiments.
ISSN:0263-2241
1873-412X
DOI:10.1016/j.measurement.2009.12.029