Catheter-aided extraction of a steel coil accidentally lodged in the right ventricle

In one case a steel coil similar to the Gianturco coil accidentally lodged in the outflow tract of the right ventricle during the embolization of a pulmonary arteriovenous fistula. Extraction of the coil with an intravascular foreign body retrieval set (Cook Inc., Bloomington Ind.) was unsuccessful...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Cardiovascular and interventional radiology 1980-01, Vol.3 (3), p.153-155
Hauptverfasser: Radojković, S, Kamenica, S, Jasović, M, Draganić, M
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In one case a steel coil similar to the Gianturco coil accidentally lodged in the outflow tract of the right ventricle during the embolization of a pulmonary arteriovenous fistula. Extraction of the coil with an intravascular foreign body retrieval set (Cook Inc., Bloomington Ind.) was unsuccessful because the extraction wire broke. An improvised device was then employed, using a Cordis 8 French multipurpose catheter and a single stainless steel wire designed for surgical osteosynthesis. The steel coil was successfully caught by the device and pulled out via the femoral vein.
ISSN:0174-1551
1432-086X
DOI:10.1007/BF02551978