A calibration procedure for the determination of alkaline phosphatase with the autoanalyzer

We describe a kinetic procedure using the LKB reaction rate analyser for the calibration of an AutoAnalyzer alkaline phosphatase method. Measurements were made at zero incubation time with a special lid for the thermostated tunnel of the LKB reaction rate analyser using assay conditions appropriate...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinica chimica acta 1980-01, Vol.105 (3), p.361-366
Hauptverfasser: Nisbet, Jennifer A., Cleeve, H.J.W., Tua, Diane
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We describe a kinetic procedure using the LKB reaction rate analyser for the calibration of an AutoAnalyzer alkaline phosphatase method. Measurements were made at zero incubation time with a special lid for the thermostated tunnel of the LKB reaction rate analyser using assay conditions appropriate to the AutoAnalyzer method. The molar absorptivity of p-nitrophenol was estimated on the LKB reaction rate analyser used for the assay of the enzyme standards. Other variables affecting the accuracy of the measurement were also determined.
ISSN:0009-8981
1873-3492
DOI:10.1016/0009-8981(80)90116-3