Internal fixation plates after fifty-six years of implantation: report of a case
Lane plates removed and examined after 56 years in the femur of a 77-year-old man were markedly corroded. There was no loss of tensile strength, but other structural properties were altered and inflammation may have been locally induced by leaching of metal ions from the steel alloy.
Gespeichert in:
Veröffentlicht in: | Clinical orthopaedics and related research 1980-06, Vol.149 (149), p.201-206 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Lane plates removed and examined after 56 years in the femur of a 77-year-old man were markedly corroded. There was no loss of tensile strength, but other structural properties were altered and inflammation may have been locally induced by leaching of metal ions from the steel alloy. |
---|---|
ISSN: | 0009-921X |