An Assessment of the Corning Electrophoretic Procedure for High Density Lipoprotein Cholesterol Estimation

The electrophoretic procedure for high density lipoprotein cholesterol determination recently introduced by Corning Medical has been assessed. The results correlated well with those obtained by phosphotungstate/Mg precipitation and ultracentrifuge procedures, but the between-batch CV was 6% in the c...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annals of clinical biochemistry 1980-05, Vol.17 (3), p.148-152
Hauptverfasser: Bullock, D G, Carter, T J N, Cosgrove, Carmel M, Hughes, Susan V
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The electrophoretic procedure for high density lipoprotein cholesterol determination recently introduced by Corning Medical has been assessed. The results correlated well with those obtained by phosphotungstate/Mg precipitation and ultracentrifuge procedures, but the between-batch CV was 6% in the clinically important range below 1 mmol/l. The densitometer, which must be carefully aligned before use for this analysis, seemed sensitive to variations in the plates and patterns produced. The technique required great care, especially with timing, and appeared expensive.
ISSN:0004-5632
1758-1001
DOI:10.1177/000456328001700309