An Improved Continuous Flow Method for Serum Creatinine Using the Jaffé Reaction

A method for the measurement of creatinine is described in which serum is dialysed directly into a combined alkaline picrate reagent. Baseline noise and drift are eliminated, and precision is improved by avoiding the need to add further reagents after dialysis. This may be of particular interest to...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annals of clinical biochemistry 1980-05, Vol.17 (3), p.153-154
Hauptverfasser: White, W A, Attwood, E C
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A method for the measurement of creatinine is described in which serum is dialysed directly into a combined alkaline picrate reagent. Baseline noise and drift are eliminated, and precision is improved by avoiding the need to add further reagents after dialysis. This may be of particular interest to laboratories that perform this assay on first-generation AutoAnalyzer equipment.
ISSN:0004-5632
1758-1001
DOI:10.1177/000456328001700310