Impaction of intra-aortic balloon due to dislocation of central stylet
A case of intra-aortic balloon rupture and impaction necessitating thoracotomy and removal through the thoracic aorta is described. Impaction was due to central stylet dislocation, producing a fish-hook effect, and was easily diagnosed from plain abdominal radiographs. A possible mechanism for style...
Gespeichert in:
Veröffentlicht in: | The Journal of thoracic and cardiovascular surgery 1980-08, Vol.80 (2), p.228-232 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A case of intra-aortic balloon rupture and impaction necessitating thoracotomy and removal through the thoracic aorta is described. Impaction was due to central stylet dislocation, producing a fish-hook effect, and was easily diagnosed from plain abdominal radiographs. A possible mechanism for stylet dislocation is suggested. |
---|---|
ISSN: | 0022-5223 1097-685X |
DOI: | 10.1016/s0022-5223(19)37795-5 |